Read this before putting the mask on anyone.

⚠️

PiEEG XR is a research device, not a medical device, and has not been certified by any regulator for use with the human body. You use it at your own risk.

Rules

  1. Battery only. Power the BCI box from a power bank or a battery head strap. Never from a wall charger, and never through a cable to a computer that is on mains power.
  2. No diagnosis. Attention, relaxation and expression values are research indicators. Don't use them to diagnose, monitor or treat any condition, or as a measure of mood or ability.
  3. Implanted devices. Don't use it on a person with a pacemaker or other implanted electronic device without a physician's approval.
  4. Skin. Use clean, intact skin. Stop if the skin becomes irritated, and clean off any gel after use.
  5. VR safety. Follow your headset maker's safety guidance. Clear the play area, and take breaks.
  6. Flashing. Update firmware with the mask off the face.

Power while worn

Power sourceWhile wornWhy
Power bank or battery head strap✅ SafeNo path to the mains.
Headset's own battery (headset not charging)✅ SafeStill fully on battery.
Wall charger❌ NeverConnects the person to the mains through the electrodes.
USB cable to a PC on mains power❌ NeverSame reason.
Headset plugged in to charge❌ Not while worn with PiEEG XRThe charger links the mains to the headset and the mask.

Hygiene

  • Wipe the electrodes and the silicone with an alcohol pad before and after each user.
  • Let them dry fully before the next use.

See also the general PiEEG safety and liabilities pages.