Electrode Placement
EasyEXG uses three electrodes: two measuring electrodes (IN+ and IN−) and one reference (REF). Each animation below shows where to place them and the signal you should see.
Electrode colours in the pictures: IN+ red · IN− yellow · REF green.
Summary
| Signal | JP1 | IN+ | IN− | REF | Sample rate |
|---|---|---|---|---|---|
| EMG (forearm) | Open · 221× | Forearm muscle, a hand-width from the wrist | Same muscle, 2–3 cm further along | Elbow | 500–1000 Hz |
| ECG (Lead I) | Bridged · 1 048× | Left chest, below the collarbone | Right chest, below the collarbone | Right lower abdomen | 250–500 Hz |
| EOG horizontal | Open · 221× | Outer corner of the right eye | Outer corner of the left eye | Centre of the forehead | 100–250 Hz |
| EOG vertical | Open · 221× | Just above one eyebrow | Just below the same eye | Forehead, to the side | 100–250 Hz |
| EEG (Fp1) | Bridged · 1 048× | Forehead above the left eye | Left earlobe | Centre of the forehead | 250 Hz |
“Right” and “left” mean the person’s own right and left. The waveforms in the animations are illustrations of typical signals, not recordings.
Before you start
- Clean the skin under each electrode with an alcohol wipe or soap and water, and let it dry. Oil and dead skin are the main cause of noise.
- Snap the electrodes onto the cable first, then stick them on. Pressing on a stuck electrode moves it.
- Set JP1 for the signal: see Gain jumper JP1.
- Power on battery only, then plug the cable into J1.
- Sit still and wait 10–20 seconds: the 0.16 Hz high-pass needs a few seconds to settle after the electrodes go on.
Muscle (EMG)

- IN+ and IN− along the muscle fibres, 2–3 cm apart, over the belly of the muscle. REF on a bony spot away from the muscle (elbow, wrist bone).
- What you see: bursts when you clench your fist, a flat line when you relax. The stronger the contraction, the bigger the bursts.
- Other muscles: biceps, calf, jaw (masseter). Keep the same rule: two electrodes on the muscle, reference on bone.
- Tip: sample at 1000 Hz for the full EMG band (20–450 Hz). Use the envelope to switch things on and off.
Heart (ECG)

- Bridge JP1 (1 048×).
- What you see: a heartbeat about once a second: a small P wave, a sharp QRS spike and a T wave.
- Wrist alternative: IN+ on the left wrist, IN− on the right wrist, REF on the right ankle or the back of the right hand. The signal is smaller but easy to set up.
- Tip: breathe normally and don’t talk. Arm movement adds muscle noise.
Eye movement, horizontal (EOG)

- JP1 open (221×).
- What you see: the trace steps one way when you look left and the other way when you look right, then slowly drifts back to centre (the high-pass filter).
- Use it for: eye-controlled cursors, games, “look left / look right” switches.
Eye movement, vertical (EOG)

- JP1 open (221×).
- What you see: slow steps when you look up or down, and a sharp spike for each blink.
- Use it for: blink detection, a “blink to click” switch.
Brain (EEG)

- Bridge JP1 (1 048×).
- What you see: with the eyes closed and relaxed, a 10 Hz alpha rhythm grows. Open the eyes and it shrinks.
- EEG is tiny (10–50 µV): sit still, relax the face and jaw, and keep cables short and away from chargers.
- Tip: filter 1–40 Hz to see it, or 8–12 Hz to see only alpha. Blinks show up as large slow bumps on Fp1.
The UNO R4’s ADC can run at 14 bits: set ADC_BITS = 14 in the Wi-Fi sketch for finer steps on EEG. See Software.
Electrode tips
| Do | Don’t |
|---|---|
| Use fresh gel electrodes on clean, dry skin | Reuse dried-out electrodes |
| Push the plug fully into J1 | Let the cable hang and pull on the electrodes |
| Keep the cable short and away from power strips | Record next to a charger or a running motor |
| Put REF on a bony, neutral spot | Put REF on a muscle you are measuring |