Troubleshooting
Signal
| Problem | Fix |
|---|---|
| Flat line | Check the wiring: OUT → A0, VCC and GND connected. Push the plug fully into J1. Check the electrodes touch clean skin. |
| Signal stuck at the top or bottom | The input is too big for the gain: open JP1 (221×) for EMG / EOG. Wait 10–20 s after the electrodes go on for the high-pass to settle. |
| Resting value not near the middle | OUT should rest at VCC/2 (about 512 of 1023 on a 10-bit ADC). If not, check VCC and GND, and that J1 and J2 aren’t both used. |
| Strong 50/60 Hz hum | Check the REF electrode first: clean skin, firm contact, on a bony spot. Run on battery, move away from chargers and power strips, keep the cable short. Set the notch to your mains frequency. |
| Signal upside down | IN+ and IN− are swapped. Swap the two electrodes, or invert in software. |
| Big swings when moving | Electrode or cable movement. Sit still, tape the cable down, use fresh electrodes. |
| No ECG heartbeat | Bridge JP1 (1 048×). Check the placement: left chest, right chest, REF on the right lower abdomen. |
| No alpha on EEG | Bridge JP1. Close the eyes for at least 10 s and relax the face. Filter 8–12 Hz. Use 14-bit ADC on the UNO R4. |
| EMG bursts too small | Put IN+ and IN− over the belly of the muscle, 2–3 cm apart, along the fibres. Sample at 1000 Hz. |
| µV values wrong by about 4.7× | The Gain setting in the viewer doesn’t match JP1 (221 vs 1048). |
| ESP32 values clipped | VCC must be 3.3 V on an ESP32, and VREF = 3.3 in the viewer. |
Uploading
| Problem | Fix |
|---|---|
| No port in Tools → Port | Try another USB cable (some only charge) or another USB socket. |
| Upload fails: “port busy” | Close the Serial Monitor / Plotter and any Python viewer, then upload again. |
This sketch needs a Wi-Fi board | The Wi-Fi sketch only builds for UNO R4 WiFi, ESP32 and ESP8266. Use the USB sketch on other boards. |
Wi-Fi and viewer
| Problem | Fix |
|---|---|
| Serial Monitor shows only dots | Wrong Wi-Fi name or password, or a 5 GHz-only network. Use 2.4 GHz. |
| Board makes an “EasyEXG” network instead of joining yours | Set USE_ACCESS_POINT 0 and upload again. |
| Viewer: “No data yet” | PC and board must be on the same Wi-Fi. Guest networks often block devices from talking to each other. |
| Still no data | Windows firewall: allow Python on Private and Public networks. |
| Viewer can’t find the board | Type the Board IP from the Serial Monitor into the “Board IP” box (or --host). |
ModuleNotFoundError | Run pip install matplotlib numpy scipy again. |
| Much fewer than 500 samples/s, “lost” keeps rising | Weak Wi-Fi: move the board closer to the router or use mode 1 (own network). |
| Small steps in the signal every 50 ms | Wi-Fi radio bursts. Use fresh electrodes, push the plug in fully, power the board from a power bank. For the cleanest signal use the USB sketch. |
Is it the board or the setup? Unplug the electrode cable: OUT should sit quietly at VCC/2. Then short IN+, IN− and REF together on J2 with a jumper: the signal should stay near zero with only a little noise. If it does, the board is fine and the noise comes from the electrodes or the environment. See Noise elimination.
Still stuck?
Ask on Discord , open an issue on GitHub , or email pieeg@pieeg.com.
Last updated on