Specification

Hardware
| Revision | EXG-Front v1 (silkscreen “EasyEXG by PiEEG V1.1”) |
| Channels | 1 differential input (IN+, IN−) plus REF / driven right leg |
| Signals | EMG, ECG, EOG, EEG |
| Instrumentation amplifier | TI INA333, gain 9.9×, CMRR ≥ 100 dB |
| Op-amp | TI OPA4388 (quad): VCC/2 reference, driven right leg, high-pass + gain, low-pass |
| Total gain | 221× (JP1 open) / 1 048× (JP1 bridged) |
| Bandwidth (−3 dB) | 0.16–501 Hz |
| Low-pass | 500 Hz, 2nd-order Sallen-Key (Q = 0.73) |
| Output | 1 analog signal, rests at VCC/2 |
| Supply | 3.3 V or 5 V, absolute max 5.5 V |
| DRL current limit | 390 kΩ, about 13 µA into the body at 5 V |
| Electrode connectors | J1: 3.5 mm TRS jack (CUI SJ-3523-SMT) · J2: 3-pin 2.54 mm header |
| Output connector | J3: 4-pin 2.54 mm header (GND, VCC, OUT, VREF) |
| PCB | 46.0 × 26.0 × 1.6 mm, 2-layer FR-4, components on top only |
| Mounting | 2 × M2 holes (Ø 2.2 mm), 40.2 mm apart |
| Isolation | None. Battery operation is required |
| Licence | CC BY-NC 4.0 |
Gain and input range
| JP1 | Use | Total gain | Band (−3 dB) | Max input before clipping (5 V / 3.3 V) |
|---|---|---|---|---|
| Open | EMG, EOG | 221× | 0.16–501 Hz | 10.9 mV / 7.2 mV |
| Bridged | ECG, EEG | 1 048× | 0.16–501 Hz | 2.3 mV / 1.5 mV |
Signal chain
| Stage | Part | Function | Gain |
|---|---|---|---|
| 1 | INA333 (U1) | Instrumentation amplifier, G = 1 + 100 kΩ / 11.2 kΩ | 9.9× |
| 2 | OPA4388 (U2C) | 0.16 Hz high-pass + non-inverting gain (JP1 selects the gain) | 22× / 106× |
| 3 | OPA4388 (U2D) | 500 Hz 2nd-order Sallen-Key low-pass (Q = 0.73) | 1× |
| — | OPA4388 (U2A) | Mid-supply reference VREF = VCC/2 | — |
| — | OPA4388 (U2B) | Driven right leg (DRL), 390 kΩ current limit | — |

Frequency response

Simulated in ngspice with a 10 MHz OPA4388 model:
- ✅ Instrumentation amplifier gain is 9.9×, as designed.
- ✅ The 0.16 Hz high-pass keeps the true ECG shape.
- ✅ The 500 Hz 2nd-order low-pass covers the whole EMG band (20–450 Hz) and removes noise above it.
- ✅ The INA333 rejects mains hum with CMRR ≥ 100 dB.
- ✅ The DRL loop is stable, with at most 2 dB of peaking (around 1.3–1.5 kHz).
Design note: R8 = 10 MΩ
R8 (DRL feedback resistor) is 10 MΩ (Yageo RC0603FR-0710ML). With 1 MΩ the right-leg drive is too weak: simulated 50 Hz common mode at the input is −46.8 dB. With 10 MΩ it improves to −55.3 dB, with peaking still below +2 dB. R9 stays at 390 kΩ, which limits the current into the body to about 13 µA at 5 V.
Simulations use idealised electrode (51 kΩ ∥ 47 nF) and body (200 pF mains coupling) models, so real hum levels will differ.
Dimensions

| Dimension | Value |
|---|---|
| Board | 46.0 × 26.0 × 1.6 mm, corner radius 1.5 mm |
| With the jack nose | 48.0 × 26.0 mm |
| Mounting holes | 2 × Ø 2.2 mm (M2) at (3.0, 16.2) and (43.2, 21.8) mm from the top-left corner |
| Height: ICs / jack / pin headers | ≈ 1.2 mm / ≈ 5.0 mm / 8.5 mm above the board |
Jack and header heights are typical catalogue values; check them against your parts before designing an enclosure.
Operating notes
- OUT rests at VCC/2 with no signal.
- Sample EMG at ≥ 1 kHz, ECG / EOG at ≥ 250 Hz, EEG at 250 Hz.
- Power the board from the same voltage as your ADC range.
- Power from a battery or an isolated supply while electrodes are on a person.
Design files
| Folder | Content |
|---|---|
Schematic | EXG-Front_v1_schematic.pdf |
Gerber | Gerber files for PCB manufacturing |
3D_Model | STEP, STL and an interactive HTML model |
Arduino, SDK | Arduino sketches and the Python viewer |
Last updated on